Installation Information

Mounting

Each ISO-PUCK Medical Grade Isolation Transformer is supplied with a steel bracket for wall mounting. If you intend to mount the ISO-PUCK on a wall or other vertical surface, attach the bracket by means of the two screws supplied.

 

 

 

Insert the cord with the NEMA 15P hospital-grade plug into a 120V wall outlet. Connect your equipment using the NEMA 15R receptacle attached to the other cord.

 

Each ISO-PUCK has a built-in auto-resettable thermal protector that will disconnect the power in case of overload, exceeding the rated power of the ISO-PUCK. You will find the rated maximum power on the label attached to the back of the unit.

 

For applications in the OEM industry, the threaded inserts allow the ISO-PUCK to be readily installed in a rack or inside equipment.

 

OPERATING FREQUENCY

Please note that if your ISO-PUCK model is marked 60Hz frequency, it should not be used at 50Hz. If your ISO-PUCK is marked 50/60Hz frequencies, it will operate on either frequency.

 

LIFE EXPECTANCY

When properly installed and used your ISO-PUCK transformer has a life expectancy of ten years or more.

 

WARRANTIES AND LIABILITIES

 

Limited Warranty

Seller warrants the ISO-PUCK transformer to be free from defects in materials and workmanship and to equal or exceed applicable published ratings and specifications or in the case of custom designed models, the specification agreed upon, under the proper storage and use for a period of 60 months from the date of original shipment. All implied warranties, including merchantability and fitness for purpose is disclaimed.


Limited Liability

Seller’s entire liability and obligation to Buyer under the above warranty shall be expressly limited to the repair, replacement or crediting, as Seller may determine at its sole discretion, of any defective or nonconforming ISO-PUCK for which Buyer has given written notice to Seller within 30 days from date of receipt of such merchandise.

 

For other terms and conditions, please refer to our General Terms and Conditions of Sale.

 

CERTIFICATIONS

UL and CSA Standards, IEC and EN Directives, and CE Marking

The transformers are certified to Medical Safety Standards UL60601- and CSA STD c22.2 No. 601.1. The design guarantees maximum 50-uA leakage current, well below the legal requirement of maximum 100 uA as specified by NEC (National Electrical Code) and UL 60601-1. The 50/60Hz models are certified to IEC60601-1 and EN61558-1 with CE marking.