Input and Output voltages of the 60Hz models are 120V. To protect against overload, each transformer has a built-in auto-resettable thermo-protector. The standard length of the 14 AWG cords is 2 feet. The 60 Hz models have a molded hospital-grade plug on the input and a molded receptacle on the output.
We have expanded our ISO-PUCK I product line to include more VA ratings. The largest size now handles 3000VA.
Download the ISO-PUCK I Specifications.pdf Download the ISO-PUCK I Brochure.pdf Download the ISO-PUCK I Price List.pdf
ISO-PUCK I 60 Hz Operating Frequency
|
Power Rating |
Model |
Diameter |
Depth |
Weight |
Inserts | ||||
|
In. |
mm |
In. |
mm |
lbs |
Kg |
Thread# |
Distance | ||
|
300 VA |
5 1/8 |
130 |
2 3/4 |
70 |
8.5 |
3.9 |
10-24 |
1.00” | |
|
400 VA |
5 1/8 |
130 |
2 3/4 |
70 |
8.5 |
3.9 |
10-24 |
1.00” | |
|
600 VA |
6 5/8 |
168 |
2 3/4 |
70 |
13 |
5.9 |
¼ -20 |
1.25” | |
|
800 VA |
6 5/8 |
168 |
2 3/4 |
70 |
13 |
5.9 |
¼ -20 |
1.25” | |
|
1000 VA |
8 1/4 |
210 |
3 |
76 |
20 |
9.1 |
¼ -20 |
1.50” | |
|
1200 VA |
8 1/4 |
210 |
3 |
76 |
22 |
10 |
¼ -20 |
1.50” | |
|
1500 VA |
8 1/4 |
210 |
3 |
76 |
25 |
11.5 |
¼ -20 |
1.50” | |
|
2200 VA |
8 1/4 |
210 |
4 |
102 |
32 |
14.5 |
¼ -20 |
1.50” | |
|
3000 VA |
9 |
230 |
4 1/4 |
108 |
38 |
17.3 |
¼ -20 |
1.50” | |
Mounting
For room installation, the ISO-PUCK I transformer is supplied with a bracket for easy single screw wall mounting. The bracket is made of heavy gauge steel and attached to the rear surface with two bolts via threaded inserts firmly anchored in the epoxy potting resin.
For applications in the OEM industry, the threaded inserts allow the ISO-PUCK I to be readily installed in a rack or inside equipment.
Certifications
UL and CSA Standards, IEC and EN Directives, CE Marking
The transformers are certified to Medical Safety Standards UL60601- and CSA STD c22.2 No. 601.1. The design guarantees maximum 50-uA leakage current, well below the legal requirement of maximum 100 uA as specified by NEC (National Electrical Code) and UL 60601-1. CB certification.
Product Features ISO-PUCK I series for 50/60Hz
For overseas markets we have new 50/60Hz designs available up to 2500 VA. For the 50/60Hz models, the input voltage is 230/240V and output voltage can be either 117V or 230/240V. In view of the different types of plugs and receptacles used outside of North America, we will be happy to hand-mount customer specified plugs and/or receptacle – or supply the cords with no termination. Custom lengths of cords are also available options.
|
Download the ISO-PUCK I specifications.pdf Download the ISO-PUCK I brochure.pdf Download the ISO-PUCK I Price List.pdf |
ISO-PUCK I 50/60 Hz Operating Frequency
|
Power Rating |
Model |
Diameter |
Depth |
Weight |
Inserts | ||||
|
In. |
mm |
In. |
mm |
Lbs |
Kg |
Thread# |
Distance | ||
|
350 VA |
5 1/8 |
130 |
2 3/4 |
70 |
8.5 |
3.9 |
10-24 |
1.00” | |
|
650 VA |
6 5/8 |
168 |
2 3/4 |
70 |
15 |
6.8 |
10-24 |
1.00” | |
|
1000 VA |
8 1/4 |
210 |
3 |
76 |
22 |
10 |
¼-20 |
1.50” | |
|
1200 VA |
8 1/4 |
210 |
3 |
76 |
25 |
11.5 |
¼-20 |
1.50” | |
|
1800 VA |
8 1/4 |
210 |
4 |
102 |
32 |
14.5 |
¼-20 |
1.50” | |
|
2500 VA |
9 |
230 |
4 1/4 |
108 |
38 |
17.3 |
¼-20 |
1.50 | |
Mounting
For room installation, the ISO-PUCK I transformer is supplied with a bracket for easy single screw wall mounting. The bracket is made of heavy gauge steel and attached to the rear surface with two bolts via threaded inserts firmly anchored in the epoxy potting resin.
For applications in the OEM industry, the threaded inserts allow the ISO-PUCK I to be readily installed in a rack or inside equipment.
Certifications
UL and CSA Standards, IEC and EN Directives, and CE Marking
The ISO-PUCK I transformers are certified to Medical Safety Standards UL60601- and CSA STD c22.2 No. 601.1. The design guarantees maximum 50-uA leakage current, well below the legal requirement of maximum 100 uA as specified by NEC (National Electrical Code) and UL 60601-1.
The 50/60Hz models are certified to IEC60601-1 and EN61558-1. CB certification. CE marking.
Visit the ISO-PUCK store to order.